144
U1401B User’s and Service Guide
8
Specifications
General Specifications
Display
•
Both primary and secondary displays are 5-digit liquid crystal display (LCD)
with a maximum reading of 51,000 counts and automatic polarity indication.
Power Consumption
•
Charging battery: 9.3 VA typical
•
DC constant current at 25 mA, maximum load: 5.5 VA typical (on 24 V DC adapter) or
2.4 VA typical (on 9.6 V batteries)
•
Meter only: 1.8 VA typical (on 24 V DC adapter) or 0.6 VA typical (on 9.6 V batteries)
Power Supply
•
Rechargeable batteries — 1.2 V × 8 pieces (Ni-MH), no cadmium, lead or mercury.
•
External switching adapter, AC 100 V to 240V, 50/60 Hz input and DC 24 V/2.5 A output.
Operating Environment
•
Full accuracy at 0 °C to 40 °C (32 °F to 104 °F)
•
Full accuracy up to 80% Relative Humidity (RH) for temperature up to 31
°
C, decreasing
linearly to 50% RH at 40
°
C
Storage Compliance
•
–20 °C to 60 °C (–4 °F to 140 °F) with batteries removed.
Safety Compliance
•
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
•
Canada: CAN/CSA-C22.2 No. 61010-1-04
•
USA: ANSI/UL 61010-1:2004
Measurement Category
•
CAT-II 150V, Pollution Degree 2 Environment.
EMC Compliance
•
IEC61326-2-1:2005/EN61326-2-1:2006
•
Canada: ICES/NMB-001:2004
•
Australia/New Zealand: AS/NZS CISPR11:2004
Measurement
•
3 times per second (AC+DC: 1 time per second)
•
1 time per second for frequency or duty cycle measurement. (>1 Hz)
•
0.25 to 1 time per second for Pulse Width measurements. (>1 Hz)
Common Mode Rejection Ratio (CMRR)
•
> 90 dB at DC, 50/60 Hz ± 0.1% (1 k
Ω
unbalanced)
Normal Mode Rejection Ratio (NMRR)
•
> 60 dB at DC, 50/60 Hz ± 0.1%
Temperature Coefficient
•
Input: 0.15 * (specified accuracy)/
°
C (from 0
°
C to 18
°
C or 28
°
C to 40
°
C)
•
Output: ± (50ppm 0.5dgt)/
°
C
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...