Performance Tests and Calibration
7
U1401B User’s and Service Guide
141
b.
Turn the rotary switch to
mV
.
Press
to select AC.
50 mV @ 1 kHz
5 mV
5 mV
5.000 mV
45 mV
45 mV
45.000 mV
500 mV @ 1 kHz
50 mV
50 mV
50.00 mV
450 mV
450 mV
450.00 mV
Turn the rotary switch to
. Press
to select
AC.
5 V @ 1 kHz
0.5 V
0.5 V
0.5000 V
4.5 V
4.5 V
4.5000 V
50 V @ 1 kHz
5 V
5 V
5.000 V
45 V
45 V
45.000 V
500 V @ 1 kHz
25 V
25 V
25.00 V
250 V
250 V
250.00 V
750 V @ 1 kHz
75 V
75 V
75.0 V
750 V
750 V
750.0 V
Table 7-5
Calibration tests
Step
Test function
Cal range
Input
U1401B
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...