Calibrator Output Operations
2
U1401B User’s and Service Guide
37
4
After selecting the required
function, press
or
to select one of two modes: Continuous or Cycle. The
secondary display will indicate
Cont
or
CyCLE
, respectively
(
Figure 2- 4
on page 38).
•
Continuous mode (Cont):
In this mode, the ramp signal is
repeated continuously. The signal will be generated
according to the amplitudes and number of steps
defined in the memory, with each step taking
approximately 0.33 seconds. For instance, according to
the default settings (
Table 2- 2
), the step size
of the
positive slope is (end amplitude – start
amplitude)/number of steps. Therefore, the step size is
(1.5 V – (–1.5 V))/15 steps = 0.2 V for
±1.5000 V
.
The step size of the negative slope is (start amplitude –
end amplitude)/number of steps. Therefore, the step
size is (–1.5 V – 1.5 V)/15 steps = –0.2 V for
±1.5000 V
.
•
Cycle mode (CyCLE):
In this mode, only one cycle of the
ramp signal is generated. The signal will be generated
according to the amplitudes and number of steps
defined in the memory, with each step taking
approximately 0.33 seconds, and then the output
amplitude will be maintained at the final value of the
ramp signal.
5
Press
OUTPUT
to start the source output. The
annunciator will appear on the display.
Table 2-2
Default settings for the autoramp output
Mode
±1.5000 V
±15.000 V
±25.000 mA
Position
Amplitude
Resolution
Amplitude
Resolution
Amplitude
Resolution
Start
–1.5000 V
015 steps
–15.000 V
015 steps
–25.000 mA
025 steps
End
+
1.5000 V
015 steps
+
15.000 V
015 steps
+
25.000 mA
025 steps
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...