Making Measurements
3
U1401B User’s and Service Guide
63
1 ms peak hold
This function allows the measurement of peak voltage for
analysis of components such as power distribution
transformers and power factor correction capacitors. The
peak voltage obtained can be used to determine the crest
factor.
Crest factor = Peak value/True- rms value
To measure the half- cycle peak voltage:
1
Press
for more than one second to toggle 1 ms peak
hold mode on or off.
2
Press
to show the peak
+
or peak– value after
activating the peak mode. The
DH MAX
annunciator
indicates the peak+ value while the
DH MIN
annunciator
indicates the peak– value. See
Figure 3- 10
on page 64.
3
If the reading is
OL
, press
to change the
measurement range and restart the peak value
measurement.
4
While the peak hold mode is on, you can press
at
any time to restart the peak value measurement.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...