U1401B User’s and Service Guide
3
Making Measurements
Percentage scale of DC mA measurement
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA
is calculated based on the measured DC mA value.
1
Select the required range (4 mA to 20 mA or 0 mA to
20 mA) in the Setup mode (refer to
Chapter 4
, “Setting
the percentage scale readout”).
2
Turn the rotary switch to
.
3
Press
to select percentage scale display for DC mA
measurement.
4
Connect the red and black test leads to the positive and
negative input terminals respectively.
5
Probe the test points in series with the circuit and read
the display. The inset display in
Figure 3- 3
shows the
percentage scale reading representing 20 mA in the range
of 4 mA to 20 mA.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...