Performance Tests and Calibration
7
U1401B User’s and Service Guide
129
Closed-case electronic calibration
This instrument features closed- case electronic calibration.
In other words, no internal electro- mechanical adjustment is
required. This instrument calculates correction factors based
on the input reference signals you feed into it during the
calibration process. The new correction factors are stored in
non- volatile EEPROM memory until the next calibration
(adjustment) is performed. The contents of the non- volatile
EEPROM memory will not change even when the power is
switched off.
Agilent Technologies’ calibration services
When your instrument is due for calibration, contact your
local Agilent Service Center to inquire about recalibration
services.
Calibration interval
A one- year calibration interval is adequate for most
applications. Accuracy specifications are warranted only if
calibration is performed at regular intervals. Accuracy
specifications are not warranted beyond the one- year
calibration interval. Agilent does not recommend extending
calibration intervals beyond 2 years for any application.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...