Installing 8890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using a Standard Column Nut
44
7000/7010 Series TQ GC/MS Operating Manual
9
Inspect the end for jagged edges or burrs. If the break is not clean and even,
repeat steps 6 and 7.
10
Adjust the column so it extends this specified distance from the end of the
transfer line. (See
For an EI XTR or CI Source Installation
on page 39.), the
column extends about 1 to 2 mm.
For an EI HES Installation
on page 39.), the column extends
about 4 to 5 mm.
Use the flashlight and magnifying loupe if necessary to see the end of the
column inside the analyzer chamber. Do not use your finger to feel for the
column end.
11
Tighten the nut 1/4 to 1/2 turn. Ensure the position of the column does not
change as you tighten the nut. Do not over tighten the nut.
12
Check the GC oven to be sure that the column does not touch the oven walls.
13
Check the nut’s tightness after one or two heat cycles; retighten as
appropriate.
14
Install the tip seal, spring, and knurled tip seal retainer on the GC/MS
interface. Align, gently slide, and screw the knurled tip seal retainer into the tip
seal holder. (See
15
Gently
check the alignment of the ion source and the interface tip seal. When
the ion source is aligned correctly, the front analyzer chamber can be closed
all the way with no resistance except the spring tension from the interface tip
seal.
16
You can align the ion source and interface tip seal by wiggling the side plate
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
17
Close the front analyzer chamber. (See
CAUTION
Use care when placing the tip seal on the end of the GC/MS interface to avoid
damaging the column.
CAUTION
Forcing the analyzer door closed if these parts are misaligned will damage the
tip seal or the interface or the ion source, or will keep the side plate from
sealing.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...