Operating in CI Mode
To Pumpdown and Switch from EI to CI Mode
7000/7010 Series TQ GC/MS Operating Manual
145
4
Select
CI/JetClean - EI/CI system
since this controller is installed for the CI
control option.
5
For
Channel A gas
, select
Methane
. Click
OK
.
6
Click the MS Tune icon in the
Instrument Control
panel to display the
Triple
Quadrupole MS Tune
window, and select the
Vacuum Control
tab to monitor
that the pressure is decreasing.
7
In the
Triple Quadrupole MS Tune
window, select the
Manual Tune
tab; then
select the
Ion Source: CI+
tab, and then in the
CI Reagent Gas Control
area
select
Gas Valve A
for
Methane
gas. (See
8
Set
Gas A (methane)
to 20%.
9
Set the GC/MS interface to 280 °C.
10
Let the system bake out and purge for at least 2 hours. If you will be running
NCI, for best sensitivity, bake out the MS overnight.
11
Click the
MS Tune
icon in the
Instrument Control
panel to display the
Triple
Quadrupole MS Tune
window and select the
Autotune
tab. The method
selects the correct PCI or NCI source and reagent gas setting.
12
Select the
Tune
from default settings
check box because you have changed
the ion source.
13
Select the
Print autotune report
check box to automatically print a tune
report.
Figure 46.
Setting reagent gas flow
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...