8
CI Maintenance
To Switch from the CI Source to the EI XTR Source
216
7000/7010 Series TQ GC/MS Operating Manual
To Switch from the CI Source to the EI XTR
Source
Procedure
1
Vent the MS. (See
on page 124.) The software will prompt
you for the appropriate actions.
2
Open the left side access panel. (See
“To Open Left Side Panel for Access to
3
Open the front analyzer chamber. (See
4
“To Remove the 7000 Series CI Source”
5
Install the EI XTR source. (See
“To Install the EI XTR Source”
6
Place the CI source in the storage container.
7
Pumpdown the MS. (See
CAUTION
Always wear clean gloves while touching the analyzer or any other parts that
go inside the analyzer chamber.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap, and take other antistatic precautions
before
you open the
analyzer chamber. (See