6
Operating in CI Mode
CI Autotune
136
7000/7010 Series TQ GC/MS Operating Manual
Front quad temp
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
Rear quad temp
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
Interface temp
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
Autotune
Yes
Yes
No
Yes
No
Yes
Table 11 CI tune default settings (continued)
Parameter
Methane
Isobutane
Ammonia