5
Operating in EI Mode
To Save a Method on the 7890 GC
7000/7010 Series TQ GC/MS Operating Manual
127
To Save a Method on the 7890 GC
Procedure
1
Press
[
Method
],
and scroll to the specific method number.
2
Press
[
Store
]
and
[
On/Yes
]
to store the new method using the chosen
number. Alternatively, press
[
Off/No
]
to return to the stored methods list
without saving the method.
A message is displayed if a method with the number you selected already
exists.
3
Press
[
On/Yes
]
to
replace
the existing method or
[
Off/No
]
to return to the
stored methods list without saving the method.