General Maintenance
To Connect or Disconnect Wiring to the EI HES
7000/7010 Series TQ GC/MS Operating Manual
161
To Connect or Disconnect Wiring to the EI
HES
This topic is used to identify the wiring connections to the ion source. (For
detailed procedures and videos, see
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Pliers, long-nose (8710-1094)
•
Tweezers (8710-2460)
Procedure
Use tweezers or pliers to connect or disconnect the ceramic board wire leads
(red, white, black, and gray) at the source connectors. (See
Figure 49.
Wiring for the EI HES
Repeller
Filament common
Filament 1
Filament 2
Ceramic source board