Installing 8890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using the Self-Tightening Column Nut
40
7000/7010 Series TQ GC/MS Operating Manual
11
Hand-tighten the nut. (See
.) Ensure the position of the column does
not change as you tighten the nut.
12
Tighten the nut in the clockwise direction. Continue to tighten until you feel
the ferrule grip the column.
13
Check the GC oven to ensure that the column does not touch the oven walls.
14
Install the tip seal, spring, and knurled tip seal retainer on the GC/MS
interface. Align, gently slide, and screw the knurled tip seal retainer into the tip
seal holder. (See
Figure 6.
Installing a capillary column in the GC/MS interface
Self-tightening
column nut
GC/MS
interface
GC inlet
Self-tightening
inlet nut
CAUTION
Use care when placing the tip seal on the end of the GC/MS interface to avoid
damaging the column.
CAUTION
Forcing the analyzer door closed if these parts are misaligned will damage the
tip seal or the interface or the ion source, or will keep the side plate from
sealing.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...