5
Operating in EI Mode
To Configure CC Gas
112
7000/7010 Series TQ GC/MS Operating Manual
To Configure CC Gas
1
Configure the use of
Quench Gas
and
Collision Gas
2
Select
Instrument > GC Parameters
from the
Instrument Control panel.
3
Click the
Configuration
icon, then select the
Modules
tab to display the
screen. (See
.)
4
From the
Collision Cell EPC >Quench Gas
drop-down menu, select the
quench gas.
5
From the
Collision Cell EPC > Collision Gas
drop-down menu, select the
collision gas.
6
Click
OK
to save the configuration.
Figure 34.
Configure CC gas