8
CI Maintenance
To Switch from the EI HES to the CI Source
222
7000/7010 Series TQ GC/MS Operating Manual
To Switch from the EI HES to the CI Source
Procedure
1
Vent the MS. (See
on page 124.) The software will prompt
you for the appropriate actions.
2
Open the left side access panel. (See
“To Open Left Side Panel for Access to
3
Open the analyzer chamber. (See
“To Open the Front Analyzer Chamber”
4
Remove the EI HES. (See
5
Place the EI HES in the storage container.
6
Remove the EI HES radiator. (See
“To Remove the EI HES Radiator”
7
Place the EI HES radiator in the storage container.
8
Remove the CI source radiator from the storage container.
9
Install the CI source radiator. (See
“To Install the CI/EI XTR Source
10
Remove CI source from the storage container.
11
Install the CI source. This requires shortening the column so that it protrudes
from the transfer line by 1 to 2 mm. (See
“To Install the 7010 Series CI
12
Pumpdown and switch from EI to CI mode. (See
CAUTION
Always verify GC/MS performance in EI before switching to CI operation.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap. (See
on page 151.) Take
antistatic precautions
before
you open the analyzer chamber.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...