5
Operating in EI Mode
To Pumpdown the MS
7000/7010 Series TQ GC/MS Operating Manual
123
16
When prompted, turn on the GC/MS interface heater and GC oven. Click
OK
when you have done so. The software will turn on the ion source and mass
filter (quad) heaters. The temperature setpoints are stored in the current tune
file.
17
After the message
Okay to run
appears, wait 2 hours for the MS to reach
thermal equilibrium. Data acquired before the MS has reached thermal
equilibrium may not be reproducible.
18
“To Autotune the MS for EI Mode”
19
If using the JetClean system, prepare the hydrogen system for use. Refer to
the
JetClean Operating manual
for warnings regarding hydrogen safety.
CAUTION
Within 10 to 15 minutes, the turbo pump speed should be up to 80% . The
pump speed should eventually reach 95%. If these conditions are not met, the
MS electronics will shut off the foreline pump. To recover from this condition,
you must power cycle the MS. If the MS does not pumpdown correctly, refer to
the manual or online help for information on troubleshooting air leaks and
other vacuum problems.
CAUTION
Do not turn on any GC heated zones until carrier gas flow is on. Heating a
column with no carrier gas flow will damage the column.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...