6
Operating in CI Mode
To Operate the Reagent Gas Flow Control Module
7000/7010 Series TQ GC/MS Operating Manual
139
To Operate the Reagent Gas Flow Control
Module
Procedure
1
In
Instrument Control
panel, select the
MS Tune
icon to display the
Triple
Quadrupole MS Tune
window. Select the
Manual Tune
tab, then select the
Ion Source
tab to display the ion source parameters. (See
.)
Figure 44.
CI flow control