8
CI Maintenance
To Install the 7010 Series CI Source
7000/7010 Series TQ GC/MS Operating Manual
211
To Install the 7010 Series CI Source
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
Slide the CI source into the radiator.
4
Install the thumbscrews. (See
5
Connect the wiring to the CI source. (See
“To Connect or Disconnect 7010
6
Close the analyzer door. (See
“To Close the Front Analyzer Chamber”
7
Pumpdown the MS. (See
8
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap, and take other antistatic precautions
before
you open the
analyzer chamber.
Figure 70.
Installing the CI source
Ion source
Thumbscrews
Source radiator