7
General Maintenance
To Close the Front Analyzer Chamber
7000/7010 Series TQ GC/MS Operating Manual
195
To Close the Front Analyzer Chamber
Procedure
1
Check the side plate O-ring.
Ensure the O-ring has a
very
light coat of Apiezon L high vacuum grease. If
the O-ring is very dry, it may not seal well. If the O-ring looks shiny, it has too
much grease on it. Refer to the
7000/7010 Series MS Troubleshooting and
2
Swing the front analyzer side plate closed.
The post-filter on the exit side of the quad helps to position the CC when the
analyzer door is closed. When closing, the door should give minimal
resistance while the quad is reseating the CC. The analyzer should slide
into place with minimal pressure.
3
Ensure the rear analyzer door is closed.
4
Ensure the vent valve is closed.
5
If hydrogen or an other flammable or toxic substance is used for carrier gas,
gently
hand-tighten the top thumbscrew on the front analyzer side plate.
6
Pumpdown the MS. (See
7
Once the MS has pumped down, close the left analyzer cover and replace the
window cover.
8
Tune the MS.
CAUTION
Do not force the analyzer door when closing, or you may damage the CC or
quadrupole.
WARNING
The top thumbscrew must be fastened if hydrogen (or other hazardous gas) is
being used as the GC carrier gas or if hydrogen is used for the JetClean
system. In the unlikely event of an explosion, it may prevent the side plate
from opening.
CAUTION
Do not overtighten the thumbscrew; it can cause air leaks or prevent
successful pumpdown. Do not use a screwdriver to tighten the thumbscrew.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...