5
Operating in EI Mode
To Vent the MS
124
7000/7010 Series TQ GC/MS Operating Manual
To Vent the MS
Agilent MassHunter GC/MS Acquisition software allows you to specify
parameters in a GC method that will automate and speed up the vent process if
direct communication (DCOMM) with the GC is established. You must predefine
a Fast Vent method to use this option. Refer to the MassHunter GC/MS
Acquisition software Online Help for information on how to configure a Fast Vent
method.
Procedure
1
Click the
MS Tune icon
from the
Instrument Control
panel.
2 Select the Manual Tune tab.
3
Select the
Vacuum Control
tab.
4
Click
Fast Vent
if your system supports DCOMM communication between
the GC and MS, or click
Vent
if it does not. If your system supports DCOMM,
MassHunter will load the Fast Vent method to turn off the GC oven and
interface heaters, the source and quad heaters, and the turbo pump.
If your system does not support DCOMM or if DCOMM has been disabled,
when prompted you must set the GC/MS interface heater and the GC oven
temperatures to ambient (room temperature).
5
Turn off the MS by pressing the power switch. (See
6
Unplug the MS power cord.
WARNING
If hydrogen is used as a carrier gas or JetClean system supply, the carrier and
JetClean system supply shutoff valves must be closed before turning off the
MS power. If the foreline pump is off, hydrogen will accumulate in the MS and
an explosion may occur. Read
on page 21 before operating
the MS with hydrogen gas.
CAUTION
Ensure the GC oven and the GC/MS interface are cool before turning off carrier
gas flow.
WARNING
When the MS is vented, do not put the MassHunter software into Instrument
Control view. Doing so will turn on the interface heater.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...