5
Operating in EI Mode
To Set the JetClean Parameters for the Clean Only Mode
114
7000/7010 Series TQ GC/MS Operating Manual
To Set the JetClean Parameters for the Clean
Only Mode
1
Click the
MS Parameters
icon in the
Instrument Control
view.
2
Select
Source Cleaning
from the
Triple Quadrupole MS Method Editor
navigation panel.
3
Select
Clean Only
from the
Operation
dropdown menu. (See
.)
4
Enter the parameters supplied by your method developer.
5
Save the method.
Figure 36.
JetClean set to Clean Only mode