7
General Maintenance
To Remove the EI HES
7000/7010 Series TQ GC/MS Operating Manual
159
To Remove the EI HES
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
Remove the two large thumbscrews that hold the ion source in place. (See
4
Disconnect wiring from the EI HES. Do not bend the wires any more than
necessary. (See
“To Connect or Disconnect Wiring to the EI HES”
WARNING
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Use an antistatic wrist strap, and take other antistatic precautions before
touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.