To Open Left Side Panel for Access to the
Analyzer Chambers
The left side panel should only be opened to access the front and rear analyzer
chambers or the analyzer side plates. This is necessary to pumpdown, clean or
change the ion source, replace a filament, or replace the EM horn. If you need to
open the left side panel follow this procedure. (See
.)
Procedure
Pull gently on the front left window, and allow the left side panel to slide forward
and down.
Figure 39.
The left side panel
Left side panel