Installing 8890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using a Standard Column Nut
42
7000/7010 Series TQ GC/MS Operating Manual
To Install a Capillary Column in the GC/MS
Interface Using a Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self-tightening
column nut explained in the previous section.
Materials needed
•
Interface tip seal (G3870-20542) (See
•
Tip seal spring (G7005-20024)
•
Knurled tip seal retainer (G3870-20547)
•
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
•
Flashlight
•
Magnifying loupe
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Interface column nut (05988-20066)
•
Ferrules
•
0.3 mm id, for 0.10 mm id columns (5062-3507)
•
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
•
0.5 mm id, for 0.32 mm id columns (5062-3506)
•
0.8 mm id, for 0.53 mm id columns (5062-3512)
•
Septum (may be old, used inlet septum)
•
Safety glasses
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...