Installing 8890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using the Self-Tightening Column Nut
7000/7010 Series TQ GC/MS Operating Manual
39
10
Adjust the column so it extends this specified distance from the end of the
transfer line. (See
For an EI XTR or CI Source Installation,
the column extends about 1 to 2 mm.
(See
For an EI HES Installation,
the column extends about 4 to 5 mm. (See
.)
Use the flashlight and magnifying loupe, if necessary, to see the end of the
column inside the analyzer chamber. Do not use your finger to feel for the
column end.
Figure 4.
Installing a capillary column in the GC/MS interface for an
EI XTR
or a
CI
source
Column extends 1-2 mm
Tip seal
GC/MS interface
Figure 5.
Installing a capillary column in the GC/MS interface for an
EI HES
Column extends 4-5 mm
Tip seal holder
GC/MS interface