
7
General Maintenance
Maintaining the Vacuum System
7000/7010 Series TQ GC/MS Operating Manual
153
Maintaining the Vacuum System
Periodic maintenance
Some maintenance tasks for the vacuum system must be performed
periodically. (See
on page 148.) These include:
•
Checking the foreline pump fluid (every week)
•
Checking the calibration vial(s) (every 6 months)
•
Replacing the foreline pump oil (every 6 months or as needed)
•
Tightening the foreline pump oil box screws (first oil change after installation)
•
Replacing the foreline pump (typically, every 3 years)
•
Replacing the tip seals in the optional dry scroll pump (every year)
•
Checking the system for leaks (On a monthly basis for safety, or as needed
depending upon the type of gases being used in the lab.)
Failure to perform these tasks as scheduled can result in decreased instrument
performance. It can also result in damage to your instrument.
Other procedures
Tasks such as replacing an ion vacuum gauge should be performed only when
needed. Refer the
Agilent 7000/7010 Series Triple Quad GC/MS Troubleshooting
and Maintenance Manual
and the online help in the MassHunter WorkStation
software for symptoms that indicate this type of maintenance is required.
More information is available
If you need more information about the locations or functions of vacuum system
components, refer the
Agilent 7000/7010 Series Triple Quad GC/MS
Troubleshooting and Maintenance Manual
.
Most of the procedures in this chapter are illustrated with video clips on the
Agilent GC/MS Hardware User Information & Instrument Utilities DVD.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...