Installing Intuvo 9000 GC Column
Conditioning Intuvo columns
7000/7010 Series TQ GC/MS Operating Manual
51
Conditioning Intuvo columns
An Intuvo column is checked for column bleed as part of the manufacturing
process. This eliminates the need for conditioning of an Intuvo column before it
is connected to the GC/MS interface. When first connected to the GC/MS
interface, Agilent recommends running a short column conditioning procedure.
This procedure reduces initial column bleed to an insignificant amount that won’t
adversely contaminate the ion source. (See
“To Condition an Intuvo Capillary
Tips and hints
•
Always use carrier gas that is at least 99.9995% pure.
•
Always wear clean gloves when handling a component’s click and run
connectors.
•
Always wear clean gloves when handling a gasket.
•
Always wear clean gloves when handling the 9000 GC/MS Tail.
WARNING
If you are using hydrogen as a carrier gas, the hydrogen gas flow must be off
before turning off the MS power. If the foreline pump is off, hydrogen will
accumulate in the MS and an explosion may occur. Read
on page 21 before operating the MS with hydrogen gas.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...