7
General Maintenance
To Assemble the EI HES
7000/7010 Series TQ GC/MS Operating Manual
179
5
Place the heater/sensor assembly over the guide pins on the source body
with the four electrical pins oriented down the flat side of the source body.
6
Place the repeller onto the heater/sensor assembly with the flat side of the
repeller circumference aligned with the interface socket in the source body.
7
Place the source mount over the repeller.
8
Finger tighten the two gold plated screws using a T6 Torx screwdriver to
secure the source mount to the source body.
9
Thread the finger grip into the filament block. (See
and
10
On the side of the filament block opposite the finger grip, orient the dual
filament ceramic holder so that it lines up with the flat of the filament block.
Fully insert the three filament leads into the filament block.
11
Inset the filament block into the source holder, and use a T6 Torx screwdriver
to secure it to the holder with the gold plated screw.
Figure 57.
Secure the lens screw and locking ring insulator
CAUTION
Do not overtighten the screws in the source body. Damage may occur to the
repeller.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...