7
General Maintenance
To Remove the EI HES Filaments
192
7000/7010 Series TQ GC/MS Operating Manual
To Remove the EI HES Filaments
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Cloths, clean (05980-60051)
•
T6 Torx driver, (8710-2548)
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
Remove the ion source, and place it on a clean cloth on your work surface.
(See
4
Remove the screw securing the filament block to the source mount. (See
.)
5
Use the hand grip on the filament block to remove the filament block from
the source.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
Figure 63.
Replacing the dual filament
Source mount
Screw
Filament block
Dual filament