
5
Operating in EI Mode
View the TQ parameters
7000/7010 Series TQ GC/MS Operating Manual
99
GC/MS Instrument Configuration program. See Configure the system in the
7890A+ or 7890B GC in the GC/MS Instrument Configuration program help.
9
In the GC/MS Instrument Configuration program, edit the TQ IP address
using these new parameters.
View the TQ parameters
This procedure allows you to monitor TQ parameters without running the
MassHunter GC/MS Acquisition software. It is useful when manually pumping
down the TQ and monitoring the instrument's status.
1
Press the
MS On/Off
switch to start the instrument.
2
Enter the instrument IP address into a PC web browser URL to display the
WUI Main Menu page. (See
Figure 28.
WUI main menu
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...