7
General Maintenance
To Install the EI HES
186
7000/7010 Series TQ GC/MS Operating Manual
4
Snap the source into place so that the source mount tabs are flush against
the source mount attachment surface of radiator. Some force is required to
overcome the resistance of the electrical contacts, spring loaded surfaces.
5
Connect the ion source wires. (See
“To Connect or Disconnect Wiring to the
6
Install and hand-tighten the source thumbscrews. Do not overtighten the
thumbscrews. (See
7
Close the front analyzer chamber. (See
8
Pumpdown the MS. (See
9