7
General Maintenance
To Clean the EI XTR Source
7000/7010 Series TQ GC/MS Operating Manual
175
.
Procedure
1
If the contamination is serious, such as an oil backflow into the analyzer,
seriously consider replacing the contaminated parts.
2
Abrasively clean the surfaces that contact the sample or ion beam.
Use an abrasive slurry of alumina powder and reagent-grade methanol on a
cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also
abrasively clean the discolorations where electrons from the filaments
enter the source body.
CAUTION
If insulators are dirty, clean them with a cotton swab dampened with
reagent-grade methanol. If that does not clean the insulators, replace them. Do
not abrasively or ultrasonically clean the insulators.
CAUTION
The filaments, source heater assembly, and insulators cannot be cleaned
ultrasonically. Replace these components if major contamination occurs.
Figure 55.
EI XTR source parts to be cleaned
Entrance lens
Ion focus lens
Extractor lens
Repeller block
insert
Source body
Repeller
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...