Microalbumin (MULTIGENT
®
)
Specimen Storage
If specimens cannot be analyzed immediately, they should be
stored observing the following guidelines:
•
Up to two weeks at 4°C
•
Up to five months at -70°C
•
Repeated freeze/thaw cycles should be avoided
•
Frozen specimens must be adequately mixed prior to
testing; if not, reduced values could be observed
Frequently Asked Questions (FAQ)
Question
Response
Can I use another vendor’s controls, or is use
of MULTIGENT Microalbumin controls
required?
Bovine or other sourced control material,
e.g.
,
Bio-Rad controls, do not work with the
MULTIGENT Microalbumin assay. The control
must be a human-sourced control.
Use MULTIGENT Microalbumin controls,
LN 2K98-10.
If controls are out of range, what
troubleshooting should be performed?
1. Check the dryer tip. If the dryer tip is not
perfectly aligned, residual water or excess
protein can remain in the cuvettes.
2. Inspect the sample probe for protein
build-up.