PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
126
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Cross Reactivity
The following digoxin cross reactivity data are not included in
the package insert.
NOTE:
Test results were obtained with an alternate
assay configuration with the same antibody as
MULTIGENT
®
Digoxin.
Compound
Conc. of Cross
Reactant
Spiked
(ng/mL)
Digoxin
Reported
Value
(ng/mL)
% Cross
Reactivity
Acetaminophen
1,000,000
0.0
0.0%
Cortisol
(hydrocortisone)
10,000
0.0
0.0%
Cortisone
10,000
0.0
0.0%
Dehydroisoandrosterone
10,000
0.0
0.0%
D
(+)-Digitoxose
10,000
0.0
0.0%
Estriol
10,000
0.0
0.0%
Furosemide
50,000
0.0
0.0%
Gitoxin
150
0.7
0.5%
Hydrochlorothiazide
100,000
0.0
0.0%
11
α
-Hydroxyprogesterone
10,000
0.1
0.001%
17
α
-Hydroxyprogesterone
10,000
0.0
0.0%
Lidocaine HCl
100,000
0.0
0.0%
Ouabain octahydrate
150
2.8
1.9%
Phenytoin
100,000
0.0
0.0%
Prednisolone
10,000
0.0
0.0%
Prednisone
10,000
0.0
0.0%
Procainamide
100,000
0.0
0.0%
Progesterone
10,000
0.0
0.0%
Propanolol HCl
100,000
0.0
0.0%
Quinidine sulfate
dihydrate
100,000
0.0
0.0%
Salicylate
1,000,000
0.0
0.0%
Secobarbital
100,000
0.0
0.0%
Testosterone
5,000
0.0
0.0%