Flowcharts
AEROSET
®
System Troubleshooting Guide
259
94816-107—November 2004
ICT™ Failed Calibration
Review error codes &
troubleshoot per Operations
Manual. Resolved?
Bubbles in ICT
module tubing or
1 mL syringe?
No
Determine whether Serum and/or Urine ICT failed calibration
Verify ICT module lot
number & installation
date
No
Yes
Replace ICT
module
Yes
Yes
ICT module:
1 Onboard > two months?
2 More than 15,000 samples?
3 Past expiration date?
Proceed
to Bubbles
in ICT
Flowchart
Error Log Message with
1.02ER000 software
Error Log Message with
1.02ER000 software
Remove bubbles
with applicator stick
& recalibrate. OK?
Verify calibrator positions,
target values, correct cals
used, open & exp. dates.
All OK?
Bubbles in calibrator
sample cup?
1 Verify ICT Dil lot number
& onboard date.
2 Verify cartridge location.
3 Verify reagent loadlist.
OK?
Yes
Stop
No
No
Correct &
recalibrate. OK?
Yes
Yes
No
No
Stop
No
B
No
Yes