Assays/Reagents
AEROSET
®
System Troubleshooting Guide
151
94816-107—November 2004
Methadone
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 300 ng/mL methadone (Multi DOA Cutoff calibrator)
when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
N
-Acetylprocainamide (NAPA)
400
l
-Alpha-methadol
2
Buprenorphine
100
Caffeine
1,000
Cimetidine
1,000
Clomipramine
2.5
Clonidine
1,000
Cotinine
100
Cyclobenzaprine
28
Desipramine
800
Diphenhydramine
500
Doxepin
10
2-Ethylidene-1,5-dimethyl-3,
3-diphenylpyrrolidine (EDDP)
1,000
Fluoxetine
500
Glutethimide
500
Ibuprofen
1,000
Ketamine
100
Ketorolac tromethamine
1,000
Lysergic acid diethylamide (LSD)
10 ng/mL
Naproxen
1,000
Phenytoin
1,000
Ranitidine
900
Scopolamine
500
Tramadol
100
Tyramine
100
Zidovudine (AZT)
2 mg/mL
Zolpidem
100