234
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Reagent Arm Adjustment
NOTE:
Ensure the reagent probe is not bent or
damaged before performing any of the following
adjustment procedures.
NOTE:
The following procedures can be used for both
the R1 and R2 probe reagent arm adjustment.
Materials Required
Empty reagent cartridge, large size
R1A Probe Horizontal Adjustment
NOTE:
Any robotic position can be adjusted using the
Adj
(Adjust) button, clockwise or counterclockwise,
after moving the robotic arm to that particular
position.
1. Select the
Maintenance Utilities
icon in the Action Area
of the Main Display. The
MAINTENANCE UTILITIES
screen displays.
2. Select the
Rgt Area 1
tab. The
Rgt Area 1
page displays.
3. Install a clean, empty reagent cartridge in position A1 of
the Reagent Supply Center.
Figure 18: MAINTENANCE UTILITIES Screen, Rgt Area 1
Page
A-LIne (Outer)
A <Adj>