PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
42
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Chemical Reactions for Clinical Chemistry Assays
Assay
Name
Reaction
Tips
ACP
α
-Naphthylphosphate
α
-N PO
4
-3
α
-N Diazotized 2-amino-5-chlorotoluene
Dye complex
Use fresh,
nonhemolyzed,
nonicteric, and
nonlipemic serum
samples. All
specimens need to be
stabilized with 50 µL
Liquid Stabilizer for
every 1 mL of serum.
The acidified serum
should be analyzed
immediately, or
stored at 2 to 8°C, or
frozen. Calibration is
stable for
approximately five
days and performed
using a water blank.
Sample Interference
Indices may assist
with determination of
sample integrity.
AlbG
(BCG)
A BCG
Colored
complex
BCG = Bromcresol green
Results are running
high, due to
non-specific protein
binding. Recommend
AlbP if more
specificity needed.
AlbP
(BCP)
Human A BCP
Colored complex
BCP = Bromcresol purple
Human albumin
specific.
AlkP
p
-Nitrophenylpho H
2
O
p
-Nitro phosphate
AlkP reagent contains
AMP buffer.
ALT
L
-A
α
-Ketoglutarate
Py
L
-glutamate
Py NADH
L
-l NAD
+
LD = Lactate dehydrogenase
NADH = Nicotinamide adenine dinucleotide
Do not use plasma
samples collected
using ammonium
heparin.
* MULTIGENT
®
assay
ACP
AlkP
Mg
+2
, Zn
+2
ALT
LD