7. Select <
CW
> (clockwise) or <
CCW
> (counterclockwise) as
needed to center the probe over the reagent cartridge.
NOTE:
If unable to align the reagent probes,
contact your local Abbott Representative.
8. Select
<OK>
to save the position.
9. Select
<OK>
on the
MAINTENANCE UTILITIES
screen
then select
<OK>
on the
Home All Robotics
dialog
window.
Clockwise
Counterclockwise