Assays/Reagents
AEROSET
®
System Troubleshooting Guide
93
94816-107—November 2004
Procedure
1. Prepare one blank level for each assay to be tested.
2. Select <
Database
> in the Information Access Area of the
Main Display. The
DATABASE
screen displays.
3. Select <
Order
> in the right column of the screen. The
Order Samples
screen displays. If this is not the first
sample ordered, select <
New
>.
4. Enter the following information:
•
Name or SID
•
C/P
•
Replicates (5)
5. Select the assay(s) to be evaluated.
6. Select <
Copy
> in the right column of the
Order Samples
screen. The
Copy Order
dialog window displays.
7. Enter
1
in the How Many Copies? field to request one copy
(N = 10, 2 samples × 5 replicates) then select <
OK
>. The
Order Samples
screen displays.
8. Load the sample(s) in the designated C/P.
9. Select <
OK
> in the right column of the screen.
10. Select <
RUN
> in the Action Area of the Main Display. The
RUN OPTIONS
screen displays.
11. Select the Patient option (
聺
).
12. Select <
Start
> in the right column of the
RUN OPTIONS
screen to initiate the run.