Assays/Reagents
Not for Customer Distribution
24
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
4. The
Outline
page is the default view.
Figure 1:
Assay Configuration Screen, Outline Page
5. Verify Quantitative Ranges is selected (
聺
). This is the
default setting.
6. Select the
Calibration
tab to access the
Calibration
page.
Figure 2:
Assay Configuration Screen, Calibration Page
7. Verify the Factor option is selected in the Calib Mode field.
8. Verify
1000
is entered in the Factor field.
9. Select
<OK>
in the upper-right corner of the
Assay
Configuration
screen. The
Save
dialog window displays.
10. Select
<OK>
to save the settings.
11. Repeat steps 1 through 10 for each DOA assay installed.
Select
Factor
Содержание AEROSET