Amphetamine/Methamphetamine
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 1,000 ng/mL
d
-methamphetamine (Multi DOA Cutoff
calibrator) when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen
1,000
Acetylsalicylic acid
1,000
Alpha-acetyl-
N
,
N
-dinormethadol
(dinor LAAM)
25
l
-Alpha-acetylmethodol (LAAM)
25
Buprenorphine
100
Caffeine
1,000
Clomipramine
2.5
Clonidine
1,000
Codeine
500
Cotinine
100
Cyclobenzaprine
28
Doxepin
10
Glutethimide
500
Ketamine
100
Lysergic acid diethylamide (LSD)
10 ng/mL
Meperidine
1,000
Phenytoin
1,000
Tramadol
100
Zidovudine (AZT)
2 mg/mL
Zolpidem
100