Flowcharts
AEROSET
®
System Troubleshooting Guide
255
94816-107—November 2004
QC Out of Range, Continued
1. Check for the following:
•
Involved assays on same line?
•
Involved assays using similar wavelengths?
•
Is the reagent kit make up the same for all assays
involved? R1 vs. R2?
2. Multiple assays? Check for the following:
•
Test file parameters are correct
•
Recent QC or reagent lot changes
•
Recent shifts or trends in QC data
•
Is QC material assayed or unassayed?
•
Manufacturer claims for analyte stability
•
Are ranges appropriate to assess system performance?
•
Peer group statistics
3. Verify reagent handling:
•
Prepared correctly, if preparation required
•
Reagent loaded in correct Reagent Carousel and
position
•
Expiration date
•
Date reagent was opened and placed onboard
•
Pipettes—disposable or washed?
•
Workload pattern
4. Check calibrators for the following:
•
Correct values entered
•
Dates opened
•
Placed onboard in correct positions
•
Length of time calibrators have been onboard
5. Verify instrument:
•
Check when source lamp was installed
•
Ensure all scheduled maintenance is current
•
Check for damaged probes, drips, or drops
•
Check cuvette washer performance
•
Check for instrument leaks,
e.g.
, pumps and syringes