Flowcharts
AEROSET
®
System Troubleshooting Guide
249
94816-107—November 2004
2. If issue is only on one line, troubleshoot:
•
Sample probes for damage, obstruction, and
alignment
•
Reagent probes for damage, obstruction, and
alignment
•
All probe tubing for crimps or leaks
•
Sample, reagent, and wash solution syringes
•
Mixers
•
Maintenance
•
Probe/mixer wash flow