32
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Summary of Print Options
The following table summarizes the three print options for the
DOA assays.
NOTE:
For troubleshooting purposes, Absorbance mode
is recommended.
ICT™ Configuration
I-Index—Calibrator/Control Position 42
I-Index is not currently used, but Calibrator/Control
position 42 (I-Index) must have a concentration of zero
defined for all ICT assays. If the concentrations are not
defined:
•
ICT buttons do not display on the
SELECT ASSAYS FOR
CALIBRATION
screen.
•
Calibration Summary report does not print the ICT Cal
L values.
Print Options
Description of Results
Outline Page
Configuration
Calibration Page
Configuration
Factor Mode—
Numeric Results
Positive or negative
numbers will print;
numbers are relative to
the cutoff calibrator
and do not represent
actual concentration.
Select Quantitative
Ranges
Calib Mode = Factor
Factor = 1,000
Factor Mode—
POS/NEG Results
Results will print POS
or NEG relative to the
cutoff calibrator.
Select Qualitative
Ranges
Calib Mode = Factor
Factor = 1,000
Absorbance Mode
Results will print as
absorbance values;
results should be
interpreted as positive
or negative relative to
the absorbance of the
cutoff calibrator.
Select Quantitative
Ranges
Edit the L-Reference
field to 0.0
Calib Mode = Abs