![background image](http://html1.mh-extra.com/html/abbott/aeroset/aeroset_troubleshooting-manual_450971228.webp)
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Hardware
Not for Customer Distribution
228
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Figure 15: Clot Detection Pressure Sensors (Close-up)
Fatal Error: GPIB Error Log Message X (X equals 1 through 11)
Probable Cause(s)
Corrective Action(s)
• GPIB cable is loose or defective.
• CPU printed circuit board is
defective.
• Flash memory data are corrupt.
• SCC previously configured in
Demo mode and software reloaded.
1. Reseat the GPIB cable.
2. Reset error in the Error Log.
3. Cycle power to the analyzer.
4. Flash memory.
5. Reconfigure SCC to Demo mode.
Cable to
PCB
Tubing to Probe
Sensor
Tubing to Syringe
Содержание AEROSET
Страница 19: ...AEROSET System Troubleshooting Guide 19 94816 107 November 2004 Assays Reagents Assays Reagents ...
Страница 221: ...AEROSET System Troubleshooting Guide 221 94816 107 November 2004 Section 10 Hardware Hardware ...
Страница 245: ...AEROSET System Troubleshooting Guide 245 94816 107 November 2004 Flowcharts Troubleshooting Flowcharts ...