AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
If any trends or sudden shifts in value of the verifiers occur,
check the assay reagent cartridge for onboard dating and
reagent expiration date. If necessary, replace the reagent
cartridge and repeat calibration.
NOTE:
The Opiates 300 calibrator contains
methamphetamine. This does not interfere with the
Opiates 300 assay.
Refer to the assay-specific package inserts for additional
information on calibration, verifiers, and controls.
Assay
Drug
Multi DOA
Verifier I
ng/mL
Opiates
300 Cal
ng/mL
Multi DOA
Cutoff Cal
ng/mL
Multi DOA
Verifier II
ng/mL
Amphetamine/
Methamphetamine
d
-Methamphetamine
0
300
1,000
2,000
Barbiturates
Secobarbital
0
200
800
Benzodiazepines
Lormetazepam
0
200
1,000
Cannabinoids
11-nor-
∆
9
-THC-9-
COOH
0
50
200
Cocaine
Benzoylecgonine
0
300
1,000
Methadone
Methadone
0
300
1,000
Opiates
Morphine
0
300
2,000
4,000
Phencyclidine
Phencyclidine
0
25
100
Propoxyphene
Propoxyphene
0
300
1,000