PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Hardware
Not for Customer Distribution
240
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
4. Verify the R1A probe is centered over the A-Line cuvette in
the Reaction Carousel.
5.
6. Select <
Adjust
> on the right side of the screen.
If...
Then
R1A probe is not centered
over the cuvette
Proceed to step 6.
R1A probe is centered over
the cuvette
Proceed to step 9.
A-Line
Cuvette
Содержание AEROSET
Страница 19: ...AEROSET System Troubleshooting Guide 19 94816 107 November 2004 Assays Reagents Assays Reagents ...
Страница 221: ...AEROSET System Troubleshooting Guide 221 94816 107 November 2004 Section 10 Hardware Hardware ...
Страница 245: ...AEROSET System Troubleshooting Guide 245 94816 107 November 2004 Flowcharts Troubleshooting Flowcharts ...