AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
7. Select <
CW
> (clockwise) or <
CCW
> (counterclockwise) as
needed to center the probe over the wash cup.
8. Select
<OK>
to save the position.
9. Select
Rgt Arm A Vert Down <Adj>
.
10. Verify the R1A probe is centered in the wash cup.
11.
12. Select
Rgt Arm A Vert Home/Up <Adj>
.
13. Repeat steps 3 through 12.
NOTE:
If the probe fails to align in the wash cup,
contact your Abbott Representative.
14. Select
<OK>
on the
MAINTENANCE UTILITIES
screen
then select
<OK>
on the
Home All Robotics
dialog
window.
If...
Then
Reagent probe is not
centered in the wash cup
Proceed to step 12.
Reagent probe is centered
in the wash cup
Proceed to step 14.