Absorbance Mode
When the system is in Absorbance mode, results print as
absorbance values. A result equal to or higher than the
Multi DOA Cutoff calibrator absorbance is considered positive
for the assay tested and negative for a value lower than the
Multi DOA Cutoff calibrator absorbance.
Perform the following steps to configure an assay in
Absorbance mode.
1. Select
<Assays>
in the Information Access Area of the
Main Display. The
ASSAYS
screen displays.
2. Select the desired assay button. The
ASSAY STATUS
screen displays.
3. Select
<Config>
in the right column of the screen. The
Assay Configuration
screen displays.
4. The
Outline
page is the default view.
Figure 5:
Assay Configuration Screen, Outline Page
5. Verify Quantitative Ranges is selected (
聺
). This is the
default setting.
6. Edit the L-Reference field from “-9999” to “0.0”.
Select